National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The component of the coefficient of variation of the method that arises from analytical variability, determined from the analysis of filters sampled at a dynamic test gas apparatus, was 0.8% to 1.3% for hydrogen chloride and 1.1% to 8.8% for nitric acid; and as determined from the analyses of spiked samples of hydrogen bromide, this value was 0.8% to 1.4% [3] .
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

With dilution of the calibration stock solution in eluent solution, prepare calibration working standard solutions covering the range of approximately 0.4 to 4 mg/L of chloride, bromide and nitrate. Store working standards in tightly sealed polyethylene bottles. Prepare fresh working standards weekly. 15.
Calibrate the ion chromatograph with at least six working standards covering the range of 0.4 to 4 mg/L of chloride, bromide and nitrate per sample by preparing a calibration graph of anion peak height (mm or μS [micro-siemens] ) vs. concentration (mg/L) .
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Wear gloves, lab coat, and safety glasses while handling acids. All work should be performed with adequate ventilation for personnel and equipment. It is imperative that acid be added to water in order to avoid a violent exothermic reaction. Avoid direct contact since acids are both corrosive and irritants to eyes, skin, and the respiratory system [8,9] . SAMPLING, SAMPLE TRANSPORT AND STORAGE: 1.
Prior to sampling, load each clean sampler, first with a sampling (impregnated) filter, then with a prefilter, separating the filters with a spacer.
Source: Wikisource