National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Method 7082, Issue 2, dated 15 August 1994 - Page 6 of 7 or 12 sample vessels. Any vessels containing 5 mL of nitric acid for reagent blank purposes are counted as sample vessels. When fewer than the recommended number of samples are to be digested, i.e., three samples plus one blank, the remaining vessels should be filled with 5 mL of nitric acid to achieve the full complement of vessels. This provides an energy balance since the microwave power absorbed is proportional to the total mass in the cavity [14] . ”