National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calibrate daily with at least seven working standards. a. Prepare standards in capped vials by mixing various volumes (i.e., 5 to 230 μL) of calibration stock solution with carbon disulfide to a total volume of 4.00 mL to obtain concentrations in the range of 0.010 to 5 mg/mL. b. Analyze together with samples and blanks (steps 11 and 12) . Determine the desorption efficiency (DE) at least once for each lot of charcoal used for sampling in the calibration range (step 8) . Prepare three tubes at each level plus three media blanks. a. Use the 400-mg charcoal tube for DE determination. ”