National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Determine desorption efficiency (DE) at least once for each batch of Chromosorb 102 used for sampling in the calibration range (step 8) . Prepare three tubes at each of five levels plus three media blanks. a. Remove and discard back sorbent section of a media blank sampler. b. Inject a known amount of calibration stock solution directly onto front sorbent section with a microliter syringe. c. Cap the tube. Allow to stand overnight. d. Desorb (steps 5 through 7) and analyze together with working standards (steps 11 and 12) . e. Prepare a graph of DE vs. μg Mevinphos recovered. ”
Source: Wikisource

National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Mevinphos is a cholinesterase inhibitor and can be absorbed through the skin. Wear protective gloves and suitable protective clothing when handling the pure material [1] .
SAMPLING: 1. 2. 3. 4.
Calibrate each personal sampling pump with a representative sampler in line. Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing.
”
Source: Wikisource

Get perspective with Kwize: daily news enlightened by great literature