National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ This method uses 13-mm glass fiber filters which have not been evaluated for collecting PCB. In Method S120, however, Aroclor 1242 was completely recovered from 37-mm glass fiber filters using 15 mL isooctane [8,13,14] . With 5 mL of hexane, Aroclor 1016 was also completely recovered from 100mg Florisil beds after one-day storage [1] . Thus, with no adsorption effect likely on glass fiber filters for PCB, 5 mL hexane should be adequate to completely extract PCB from combined filters and front sorbent sections. Sample stability on glass fiber filters has not been investigated. ”