National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

ESTIMATED LOD: 1 μg each analyte per sample PRECISION:
See EVALUATION OF METHOD.
APPLICABILITY: This method may be used to determine two or more of the specified analytes simultaneously.
INTERFERENCES: High humidity reduces sampling capacity. Less volatile compounds may displace more volatile compounds on the charcoal.
OTHER METHODS: This method combines and updates Methods 1401 and 1402, NMAM, Fourth Edition, Issue 2. Estimated LOD for each analyte is approximately ten times lower than that of the old methods.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Inject a known amount of stock solution directly onto front sorbent section with a microliter syringe. NIOSH Manual of Analytical Methods (NMAM) , Fourth Edition ALC OH OLS CO MB INED : MET HO D 140 5, Issue 1, dated 15 M arch 200 3 - Page 3 o f 6 c. Ca p the tube. Allow to stand overnight. d. Deso rb (steps 5 through 7 ) and analyze together with working standards (steps 11 and 12) . e. Prepare a graph of DE vs. μg analyte recovered. 10. Analyze three quality control bind spikes and three analyst spikes to insure that the calibration graph and D E graph are in control.
Source: Wikisource