National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Sampler: a. Filter. PTFE-laminated membrane filter, 2-μm pore size, 37-mm diameter (Gelman Zefluor, Membrana, Pleasantown, CA, or equivalent) , backed by a spacer (37-mm OD, 32-mm ID) cut from a cellulose support pad or SKC #225-23, in cassette filter holder. NOTE 1: If sampling is to be done in bright sunlight, use opaque or foil-wrapped cassettes to prevent sample degradation. NOTE 2: Take filters to be preweighed from the filter package and allow to equilibrate 24 h with laboratory atmosphere before taring. b. Sorbent tube, connected to filter with minimum length PVC tubing.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Normalize the total mass of PAH found to the mass of sample collected. e. Choose the solvent which gives the highest recovery of PAH of interest. Use the solvent chosen to extract the personal filter samples. 8. Extract filters. a. Add 5.0 mL of the solvent chosen in step 7 to each scintillation vial containing a filter. Start media and reagent blanks at this step. b. Cap and let stand 15 to 20 min in an ultrasonic bath. NOTE: Soxhlet extraction may be required when large amounts of highly adsorptive particulate matter (e.g., fly ash or diesel soot) are present. 9. Desorb PAH from sorbent.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The working range for B [a] P is 3 to 150 μg/m 3 for a 400-L air sample. Specific sample sets may require modification in filter extraction solvent, choice of measurement method, and measurement conditions.
INTERFERENCES: Any compound which elutes at the same GC retention time may interfere. Heat, ozone, NO 2, or UV light may cause sample degradation.
OTHER METHODS: This revises P&CAM 183 [2] . The spectrophotometric methods, P&CAM 184 and 186 [2] , have not been revised. Method 5506 (HPLC) uses the same sampling technique and is more sensitive.